End point, End point delay, End point method – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 91: Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 91

End point

Adjustable pH/mV value to which titrant is to be dispensed.

Enter in:

Edit method > Parameters menu (end point method)

Edit reagent > Calibration parameters menu (end point method)

Range available:

-9.000 pH to +23.000 pH or -2000.0 mV to +2000.0 mV

End point delay

Time during which the electrode signal is monitored after an end point
has been reached. If the electrode signal swings back, the burette
adds extra titrant to reach the end point and the count down of the
End point delay is repeated.

Enter in:

Edit method > Parameters menu (end point method)

Edit reagent > Calibration parameters menu (end point method)

Range available:

00:00 to 99:59 min:s

End point
method

For this method, the user enters one or several pH/mV values, corre-
sponding to the end points to be found during the titration.

Up to 2 end points can be determined per method.

The titration is carried out using the continuous addition technique.
The speed of addition is adjusted with respect to the distance which
separates the end point from the slope of the titration curve and the
method parameters entered by the user.