Page 169
Sample changer
To automate your entire titration procedure (up to 126 samples in one
go), the SAC80 or SAC850 sample changers can be connected to the
SAC socket of the Titration Manager using the cable, part no.
A95A202 (SAC80) or A95X501 (SAC850).
Refer to "Sample changer cable - A95A202 (SAC80)", page 170
Refer to "Sample changer cable - A95X501 (SAC850)", page 170
When using a sample changer, you have to indicate to the Titration
Manager which model is used.
Enter in:
Setup menu > Configuration
Range available:
SAC80, SAC850, No
Depending on the sample changer used, you have to enter other
configuration parameters:
•
For a SAC80:
Number of rinses,
see "Number of static rinses", page 127
,
•
For a SAC850:
Automatic detection of beakers,
see "Beaker detection", page 50
,
Number of rinses,
see "Number of static rinses", page 127
,
,
see "Sequence end in Park (Yes/No)", page
See also:
Edit a sequence with a sample changer:
Refer to "Programming sequence", page 139
Select a sequence with a sample changer:
Refer to "Select sequence", page 175
Run a sequence with a sample changer:
Refer to "Running a sequence with a sample changer", page
166
.