Sample stack, Sample unit, Sample unit conflict – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 174: See "sample, Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 174

Sample stack

The sample stack gathers individual data for all the samples present
in a sequence.

1.

Perform steps 1 to 7 of the "Programming sequence" procedure.

Refer to "Programming sequence", page 139

.

2.

In the Sequence/Sample stack window, press 1 Sample stack.

3.

Enter sample data.

Label the beakers indicating the number of beakers in the
sequence, for example: 1/11, 2/11 etc.... and the sample ID.

Place the beakers in the numbered position on the sample
changer. If static rinses and/or dynamic rinses are pro-
grammed, position the corresponding rinse beakers at the
right places.

Refer to "Number of static rinses", page 127

.

Refer to "Dynamic rinses", page 77

.

You can print the sample stack by pressing Print from the Se-
quence/Sample stack screen (screen displayed at step 2 of
the operating procedure above).

Sample unit

Unit of the sample amount. The sample unit can be expressed in
weight (g, mg) or volume (ml, µl). The sample unit indicates the result
unit available.

Enter in:

Edit method > Sample menu

Sample unit
conflict

In a Coupled method, the sample unit is not the same for the two
titration methods.

Check the Sample screens of each method of the Coupled method.

<1/11> means the first beaker over 11
programmed in the sequence.
Use the LEFT/RIGHT arrow keys to
review the other beakers in the sequence.
The data displayed will depend on the
type of method selected.

Advertising