Running a
method
(preliminary
steps)

Before running a method or a sequence (sequence with a SAC80
Sample Changer), perform the following operations in the order listed
below:

1.

Select method.

Refer to "Select method", page 175

.

2.

If Question marks "?" are present in the Reagent and/or Electrode
tabs, it means that the method needs to be programmed - a rea-
gent or an electrode is missing. Review programming in Supervi-
sor mode.

Refer to "Programming method", page 138

.

3.

Connect/Check electrodes.

Refer to "Electrode connection", page 86

.

4.

Install/Check reagent(s).

Refer to "Install reagent", page 108

.

5.

When Sunny icons are visible in the Reagent and Electrode tabs,
press 1 in the Main window.

Refer to "Reagent icons", page 149

.

Refer to "Electrode icons", page 87

.

If you have a problem running the analysis, the Titration
Manager will guide you through the necessary operations so
that you are able to run the analysis in no time at all.

Important - About the Sample Run start option :
If

Sample Run Start = Shortcut

has been

selected in the Setup > Configuration menu,
dip the electrodes into the sample before pressing 1
(step 5) !

Refer to "Sample Run Start/Run End (Standard/Shortcut)",
page 173

.

For the following steps :

Refer to "Running a method (steps 6 to 9)", page 163

.

To calibrate an electrode,

see "Electrode calibration", page 82

.

To calibrate a reagent,

see "Reagent calibration", page 145

.