Input Range Parameter

The following tables show which measurement ranges of the instrument correspond to
the input types of the CR command (control input range setting command) and SR
command (input range setting command), VOLT, TC, RTD, DI, and SQRT. The table
also shows the ranges for the upper and lower limits of the span.
DC Voltage (VOLT) and Square Root Computation (SQRT)

Measurement Range

Parameter for the

Upper and Lower

Upper and Lower Limits

CR/SR Command

Limits of the Span

of the Span for the CR/SR
Command

20 mV

20mV

–20.00 to 20.00 mV

–2000 to 2000

60 mV

60mV

–60.00 to 60.00 mV

–6000 to 6000

200 mV

200mV

–200.0 to 200.0 mV

–2000 to 2000

2 V

2V

–2.000 to 2.000 V

–2000 to 2000

6 V

6V

–6.000 to 6.000 V

–6000 to 6000

20 V

20V

–20.00 to 20.00 V

–2000 to 2000

50 V

50V

–50.00 to 50.00 V

–5000 to 5000

Thermocouple (TC)

Measurement Range

Parameter for the

Upper and Lower

Upper and Lower Limits

CR/SR Command

Limits of the Span

of the Span for the CR/SR
Command

R

R

0.0 to 1760.0 °C

0 to 17600

S

S

0.0 to 1760.0 °C

0 to 17600

B

B

0.0 to 1820.0 °C

0 to 18200

K

K

–200.0 to 1370.0 °C

–2000 to 13700

E

E

–200.0 to 800.0 °C

–2000 to 8000

J

J

–200.0 to 1100.0 °C

–2000 to 11000

T

T

–200.0 to 400.0 °C

–2000 to 4000

N

N

0.0 to 1300.0 °C

0 to 13000

W

W

0.0 to 2315.0 °C

0 to 23150

L

L

–200.0 to 900.0 °C

–2000 to 9000

U

U

–200.0 to 400.0 °C

–2000 to 4000

PLATI

PLATI

0.0 to 1400.0 °C

0 to 14000

PR

PR

0.0 to 1900.0 °C

0 to 19000

WRe

WRe

0.0 to 2400.0 °C

0 to 24000

Resistance Temperature Detector (RTD)

Measurement Range

Parameter for the

Upper and Lower

Upper and Lower Limits

CR/SR Command

Limits of the Span

of the Span for the CR/SR
Command

Pt100 PT

-200.0 to 600.0 °C

–2000 to 6000

JPt100

JPT

–200.0 to 550.0 °C

–2000 to 5500

Cu10 (GE)

*

CU1

–200.0 to 300.0 °C

–2000 to 3000

Cu10 (L&N)

*

CU2

–200.0 to 300.0 °C

–2000 to 3000

Cu10 (WEED)

*

CU3

–200.0 to 300.0 °C

–2000 to 3000

Cu10 (BAILEY)

*

CU4

–200.0 to 300.0 °C

–2000 to 3000

Cu10 α=0.00392

CU5

–200.0 to 300.0 °C

–2000 to 3000

at 20 °C

*

Cu10 α=0.00393

CU6

–200.0 to 300.0 °C

–2000 to 3000

at 20 °C

*

Cu25 α=0.00425 at 0 °C

*

CU25

–200.0 to 300.0 °C

–2000 to 3000

* Input range that can be specified on models with the Cu10, Cu25 resistance temperature detector