Xv sets the scan interval, Xt selects the temperature unit – Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 170

Advertising
background image

6-41

IM 04L31A01-17E

6

Command

XV

Sets the scan interval

Syntax

XV p1<terminator>
p1: Scan interval

1S, 2S

Query

XV?

Example

Set the scan interval to 1 s.

XV1S

XT

Selects the temperature unit

Syntax

XT p1<terminator>
p1

Temperature unit

C Celsius (°C)
F Fahrenheit (°F)

Query

XT?

Example

Set the temperature unit to Fahrenheit.

XTF

Description This command can be used on models with the

display language code “-2”.

XS

Sets the channels used to
display the trend and acquire the
data

Syntax

XS p1,p2<terminator>
p1: Channel number (01 to 20, 31 to 60, 101 to

118, 201 to 248) (01 to 06, 31 to 42, 101 to

106, 201 to 212 on the CX1000)

p2: Enable/disable (ON, OFF) displaying of the

trend and acquiring of the data

Query

XS[ p1]?

Example

Enable displaying of the trend and acquiring of

the data on channel 01.

XS01,ON

Description Computation channels (31 to 60) can only be

specified on models with the computation

function option /M1.

XM

Sets the conditions used to
acquire display/event data to the
internal memory or save the data
to the external storage medium

Syntax

XM p1,p2,p3,p4,p5,p6,p7,p8,p9,p10
<terminator>

p1: Method of saving data to the external

storage medium (AUTO, MANUAL)

p2: Data type (DISPLAY, EVENT, E+D)

p3: Sample rate of event data (125MS, 250MS,

500MS, 1S, 2S, 10S, 30S, 60S, 120S,

300S, 600S)

p4: Event mode (FREE, TRIGGER, ROTATE)

p5: Number of blocks

When p2 is set to EVENT 1, 2, 4, 8, 16

When p2 is set to E+D

1, 2, 4

p6: Event data length (3MIN, 5MIN, 10MIN,

20MIN, 30MIN, 1H, 2H, 3H, 4H, 6H, 8H,

12H, 1DAY, 2DAY, 3DAY, 5DAY, 7DAY,

10DAY, 14DAY, 31DAY)

p7: Pretrigger(0, 5, 25, 50, 75, 95, 100) [%]

p8: Turn ON/OFF manual trigger (ON, OFF)

p9: Turn ON/OFF external trigger (ON, OFF)

p10: Turn ON/OFF alarm trigger (ON, OFF)

Query

XM?

Example

Set the method of saving data to the external

storage medium to auto, the data type to both

display data and event data, the sample rate to

10 s, the event mode to TRIGGER, the event

data length to 30 minutes, the number of blocks

to 1, the pretrigger to 50%, the manual trigger to

ON, the external trigger to ON, and the alarm

trigger to ON.

XMAUTO,E+D,10S,TRIGGER,1,30MIN,50,

ON,ON,ON

Description • The setting of p6 is valid when p1 is AUTO

and p4 is FREE.

• The setting of Parameters p3 through p10 are

valid when p2 is set to EVENT or E+D.

• When p2 is set to E+D, p4 cannot be set to

FREE.

• The settings of p5 to p10 are valid when p4 is

TRIGGER or ROTATE.

• The event data length selection (p5) varies

depending on the p3 setting and the number

of channels that are measuring and

computing. For details, see the

user's manual

IM 04L31A01-01E or IM 04L31A01-03E

.

XU

Sets the channel identification
display, memory alarm time,
language, whether or not to use
the partial expanded display
function and the batch function

Syntax

XU p1,p2,p3,p4<terminator>
p1: The display used to identify the

measurement/computation channels (TAG,

CHANNEL)

p2: Memory alarm time (1H, 2H, 5H, 10H, 20H,

50H, 100H)

p3: Language (ENGLISH, JAPANESE)

p4: Use/Not use partial expanded display

function (USE, NOT)

Query

XU?

Example

Set the display used to identify the

measurement/computation channels to channel

numbers, the memory alarm length to 1 hour, the

language to English, use the partial expansion

function and the batch function.

XUCHANNEL,1H,JAPANESE,USE,USE

Description The SP command can be used when parameter

p4 of this command is set to USE.

6.7 Basic Setting Commands (Measurement)