Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 192

Advertising
background image

6-63

IM 04L31A01-17E

6

Command

6.15 Instrument Information Output Commands

6.15 Instrument Information

Output Commands
(Available when using the
instrument information
server function via
Ethernet communications)

The instrument information server function interprets one

UDP packet to be one command and returns a single packet

(containing the CX information) in response to the command.

Port number

34264/udp

(see section 2.1)

Transfer data ASCII

Received buffer length

128

Transmit buffer length

512

Maximum number of parameters

32

In the command packet, parameters corresponding to the

desired information are placed one after another.

Parameter

Description

all

Outputs all information that is output using the

parameters below (serial, model, host, ip).

serial

Outputs the serial number.

model

Outputs the manufacturer, model, and

firmware version.

host

Outputs the host name (the host name

specified in section 2.3).

ip

Outputs the IP address (the host name

specified in section 2.3).

Example

Query the IP address and host name. (Of the

two frames below, the top frame represents the

command packet, the bottom frame represents e

response packet.)

ip host

EA
ip = 192.168.111.24
host = CX2000
EN

Description • Separate each parameter with one or more

blanks (space, tab, carriage return, line feed).

• Parameters are not case sensitive.

• Undefined parameters will be ignored.

• Parameters beyond the 32nd parameter are

ignored.