Ps starts/stops measurements, Ak confirms the alarm status (alarm acknowledge) – Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 166

Advertising
background image

6-37

IM 04L31A01-17E

6

Command

p8: Display item on the lower right screen

(screen 4) (same selections as p2)

p9: Number of the group to be displayed on the

lower right screen (screen 4)

Example

Set the screen to four screen display, the display

item of screen 1 to trend display, the group

number to display in screen 1 to 1, the display

item of screen 2 to digital display, the group

number to display in screen 2 to 2, the display

item of screen 3 to bar graph display, the group

number to display in screen 3 to 3, the display

item of screen 4 to message summary display,

and the group number to display in screen 4 to 4.

UD2,TREND,1,DIGITAL,2,BAR,3,

MESSAGE,4

Description The p3, p5, p7, and p9 parameters are valid

when p2, p4, p6, and p8 are set to a display

other than OVERVIEW, respectively.

When displaying the 4 screen display set with the
SY command (CX2000 only)

Syntax

UD p1,p2<terminator>
p1: Screen switching (3)

p2: 4 screen display number (0 to 4)

0

The screen is set to the 4 screen

display that result by setting

parameter p1 to 2 using the UD

command. This setting (p1 = 3, p2 =

0) is valid only when the 4 screen

display is enabled by setting p1 to 2

beforehand.

1

Display the screen of 4 screen

display number 1 that was specified

with the SY command.

2

Display the screen of 4 screen

display number 2 that was specified

with the SY command.

3

Display the screen of 4 screen

display number 3 that was specified

with the SY command.

4

Display the screen of 4 screen

display number 4 that was specified

with the SY command.

Example

Display the screen of 4 screen display number 1

that was specified with the SY command.

UD3,1

When turning ON/OFF the automatic display
switching, switching between all channel display
and group display, enabling/disabling the scale
display, and enabling/disabling digital display
(numerical display section)

Syntax

UD p1,p2,p3,p4,p5<terminator>
p1: Screen switching (4)

p2: Enables/disables automatic display

switching (ON/, OFF)

p3: Switches between all channel display and

group display (All, Group)

p4: Turns ON/OFF the scale display

p5: Turns ON/OFF the digital display

Example

Enable the automatic display switching, switch to

the group display, turn ON the scale display, and

turn OFF the digital display.

UD4,ON,GROUP,ON,OFF

Description • Parameter p2 is valid for the trend, digital, or

bar graph displays. Use the SE command to

set the scroll interval.

• Parameters p3 and p4 are valid for the trend

display.

• Parameter p5 is valid for the trend or 4 screen

trend displays.

PS

Starts/Stops measurements

Syntax

PS p1<terminator>
p1: Starts/Stops measurements

0 Start

1 Stop

Example

Start the measurement.

PS0

Description When measurement is started, the display,

event, and report data is acquired to the internal

memory.

AK

Confirms the alarm status (alarm
acknowledge)

Syntax

AK p1<terminator>
p1: Executes alarm acknowledge (0)

Example

Confirm the current held condition of the alarm

(execute alarm acknowledge).

AK0

EV

Saves the manual sample,
manual trigger, snapshot,
display data, saves the event
data

Syntax

EV p1<terminator>
p1: Operation type

0

Execute manual sampling.

1

Activate manual trigger.

2

Take a snapshot.

3

Save display data to the external

storage medium.

4

Save event data to the external

storage medium.

Example

Execute manual sampling.

EV0

6.6 Setting Commands (Operation)