6-28

IM 04L31A01-17E

p4: Measurement range

p5: Span lower limit

p6: Span upper limit

p7: Reference channel: 01 to 20 (01 to 06 on

the CX1000)

Query

SR[ p1]?

Example

Set the range mode of channel 10 to the

difference computation between channels with

the reference channel set to 01 and set the input

type to TC. Set the range to R. Set the span

lower limit to 10.0°C and span upper limit to
100.0°C.
SR10,DELTA,TC,R,100,1000,01

Description • This command cannot be specified while

measurement/computation is in progress or

while a report is being created.

• Set parameters p4, p5, and p6 according to

the table in section 6.3.

• For parameters p5 and p6, enter a value using

5 digits or less excluding the decimal point.

The decimal position is fixed to the position

indicated in the table in section 6.3.

When setting channels to scaling

Syntax

SR p1,p2,p3,p4,p5,p6,p7,p8,p9,p10
<terminator>

p1: Channel number (01 to 20) (01 to 06 on the

CX1000)

p2: Range mode (SCALE)

p3: Input type

VOLT

DC voltage

TC

TC

RTD

RTD

DI ON/OFF input

p4: Measurement range

p5: Span lower limit

p6: Span upper limit

p7: Scale lower limit (–30000 to 30000)

p8: Scale upper limit (–30000 to 30000)

p9: Scale decimal point position (0 to 4)

p10: Unit (up to 6 characters)

Query

SR[ p1]?

Example

Convert the DC voltage measured on channel 02

to a DC current. Set the measurement range to

6 V, span lower limit to 1 V, span upper limit to 5

V, scale lower limit to 1.00 A, and scale upper

limit to 5.00 A.

SR02,SCALE,VOLT,6V,1000,5000,100,

500,2,A

Description • This command cannot be specified while

measurement/computation is in progress or

while a report is being created.

• Set parameters p4, p5, and p6 according to

the table in section 6.3.

• For parameters p5 and p6, enter a value using

5 digits or less excluding the decimal point.

The decimal position is fixed to the position

indicated in the table in section 6.3.

• For parameters p7, p8, and p9, either set all

three parameters or omit all three parameters.

When setting channels to square root
computation

Syntax

SR p1,p2,p3,p4,p5,p6,p7,p8,p9
<terminator>

p1: Channel number (01 to 20) (01 to 06 on the

CX1000)

p2: Range mode (SQRT)

p3: Measurement range

p4: Span lower limit

p5: Span upper limit

p6: Scale lower limit

(–30000 to 30000)

p7: Scale upper limit

(–30000 to 30000)

p8

Scale decimal point position (0 to 4)

p9

Unit (up to 6 characters)

Query

SR[ p1]?

Example

Convert the DC voltage measured on channel 01

to the amount of flow using the square root

computation. Set the measurement range to 6

V, span lower limit to 1 V, span upper limit to 5 V,

scale lower limit to 10.0 m

3

/s, and scale upper

limit to 100.0 m

3

/s.

SR01,SQRT,6V,1000,5000,100,1000,1,

m

3

/s

Description • This command cannot be specified while

measurement/computation is in progress or

while a report is being created.

• Set parameters p3, p4, and p5 according to

the table in section 6.3.

• For parameters p4 and p5, enter a value using

5 digits or less excluding the decimal point.

The decimal position is fixed to the position

indicated in the table in section 6.3.

• For parameters p6, p7, and p8, either set all

three parameters or omit all three parameters.

SO

Sets the computing equation

Syntax

SO p1,p2,p3,p4,p5,p6,p7<terminator>
p1

Computation channel number (31 to 60)

(31 to 42 on the CX1000)

p2

Turn ON/OFF computation (On, Off)

p3

Computing equation (up to 40 characters)

p4

Span lower limit (–9999999 to 99999999)

p5

Span upper limit (–9999999 to 99999999)

p6

Span decimal point position (0 to 4)

p7

Unit (up to 6 characters)

Query

SO[ p1]?