8.2

Bit Structure of the Status Information

The following four groups of status information are output in response to a status
information output request using the IS command. For a description of the output format,
see “Status Information” in section 7.2, “Output Format of ASCII Data.”

Status Information 1

Bit

Name

Description

0

A/D conversion complete

Set to “1” when the A/D conversion of the measurement is
complete.

1

Medium access complete

Set to “1” when the display, event, manual sampled, report,
TLOG, or screen image data file are finished being saved to
the external storage medium.
Set to “1” when setting data is successfully saved or loaded.

2

Report generation complete

Set to “1” when report generation is complete.

3

Timeout

Set to “1” when the timer expires.

4

5

6

7

Status Information 2

Bit

Name

Description

0

Measurement dropout

Set to “1” when the measurement process of the
measurement channel could not keep up.

1

Decimal point/unit information

Set to “1” when the decimal point/unit information has been

change

changed.*

2

Command error

Set to “1” when there is a command syntax error.

3

Execution error

Set to “1” when an error occurs during command execution.

4

5

6

7

* Set to “1” if there is a change in the measurement/computation/control channel.

Status Information 3

Bit

Name

Description

0

1

2

Memory end

Set to “1” while the free space in the internal memory or
external storage medium is low (see section 1.19 in the

user’s manual IM 04L31A01-01E

or

IM 04L31A01-03E

.

3

4

5

6

7

Status Information 4

Bit

Name

Description

0

Basic setting

Set to “1” during basic setting mode.

1

Memory sampling

Set to “1” while data are being acquired into the internal
memory.

2

Computing

Set to “1” while computation is in progress.

3

Alarm occurring

Set to “1” while alarm is occurring.

4

Accessing medium

Set to “1” while the display, event, manual sampled, report,
TLOG, or screen image data file are being saved to the
external storage medium.

5

E-mail started

Set to “1” while the e-mail transmission is started.

6

Controlling

Set to “1” while executing control operation.

7

Advertising