Decimal point position/unit information, Decimal point position/unit information -8 – Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 200

Advertising
background image

7-8

IM 04L31A01-17E

Decimal Point Position/Unit Information

• The FE command is used to output the data.

Syntax

The data is output for each channel in the following syntax.
EACRLF
s_cccuuuuuu,ppCRLF
..................
ENCRLF

s

Data status (N, D, S)
N: Normal
D: Differential input
S: Skip (When the measurement range is set to SKIP for a

measurement channel or when the channel is turned OFF for
a computation channel)

ccc

Channel number
0xx: Measurement channel (001 to 020) (001 to 006 on the CX1000)
Axx: Computation channel (A31 to A60) (A31 to A42 on the CX1000)
1xx: Internal control channel (101 to 118) (101 to 106 on the CX1000)
2xx: External control channel (201 to 248) (201 to 212 on the CX1000)

uuuuuu

Unit information (6 characters, left-justified)
mV____: mV
V_____: V
^C____: °C
xxxxxx: (User-defined character string)

pp

Decimal point position (00 to 04)
No decimal point (00000) for 00.
One digit to the right of the decimal (0000.0) for 01.
Two digits to the right of the decimal (000.00) for 02.
Three digits to the right of the decimal (00.000) for 03.
Four digits to the right of the decimal (0.0000) for 04.

_

Space

Example

EA
N 001mV ,01
N 002mV ,01
EN

7.2 Output Format of ASCII Data