Appendix 8 Login Procedure

You log into the CX from your PC to use the functionality of the setting/measurement
server and the maintenance/test server via the Ethernet interface. If you complete the
procedure successfully up to login complete in the following figure, the commands in
chapter 6 become functional.
When Using the Ethernet Login Function of the CX

E1 403

E1 403

Yes

No

No

No

No

No

No

E1 404

E1 421

Connect

Disconnect

E1 422

Disconnect

E1 420

Disconnect

Login complete

Within 3 retries?*

2

Within 3 retries?*

3

Wait 5 s

E1 401

E0

E1 400

Entered “quit”?

Verification match?

Within the number of
simultaneous uses?*

3

Within 2 minutes?

Within the number
of connections*

1

Within 2 minutes?

User name

Password

Increment retry count

1

1

1

1

End

Start

Judge

Transit

Code (for a description of codes and messages, see Appendix 7)

Response from the CX (message omitted)
For a description of the response format, see section 7.1.

User entry

*1 Connections cannot exceed the maximum number of connections (see section 2.1).
*2 If you try to log in using a wrong password four consecutive times, the communication is

dropped (the number of retries for login is three).

*3 If you try to log in causing the number of simultaneous uses at the administrator or user level

to be exceeded (see section 2.1) four consecutive times, the communication is dropped (even
if the password is correct).