2-3

IM 04L31A01-01E

Installation and Wiring

3

2

1

4

5

6

7

8

9

10

11

12

13

14

Installation Procedure

Use a 2 mm to 26 mm thick steel plate for the panel on which the CX2000 is to be
mounted.
1. Insert CX2000 from the front of the panel.
2. Using the mounting brackets that came with the package, insert the CX2000 to the

panel as shown in the following figure.
• Two mounting brackets are used at the top and bottom or left and right of the case

(remove the seal that is covering the holes of the mounting brackets on the case
beforehand).

• The adequate tightening torque of the screws for the panel mounting brackets is

0.7 to 0.9 N·m.

• Mount the CX2000 to the panel according to the procedure below.

• First, attach the two mounting brackets and temporarily fasten the attachment

screws.

• Next, fix the CX in place by tightening the attachment screws with the

appropriate torque. When the CX is approximately perpendicular to the panel as
you fasten the screws, press the mounting bracket against the case so that they
are in contact with each other.

CAUTION

Tightening the screws with a torque greater than the adequate tightening torque
can cause deformation of the case or damage to the bracket.

Panel Cut Diagram

361 min.

360 min.

281

+2

0

281

+2 0

Unit: mm

Panel Mounting Diagram

(The figure shows the case when the mounting

brackets are used on the top and bottom of the case.)

Screw temporarily

Fix in place

Front

Panel

Panel

Mounting bracket

Mounting bracket

Torque driver
(flat blade)

Attachment screw

Case

In contact
with each other

Attachment screw

2.2 Installation