Error messages for communication commands are output in English.

Errors Related to Parameter Settings
• Setting Errors

Code

Message

Explanation/Countermeasures/Ref. section

1

System error.

Contact your nearest YOKOGAWA dealer.

2

Incorrect date or time setting.

See section 3.3.

3

A disabled channel is selected.

See the Communication Interface User’s Manual.

4

Incorrect function parameter.

See the Communication Interface User’s Manual.

5

The input numerical value exceeds the set range.

Enter a proper value.

6

Incorrect input character string.

Enter a proper character string.

7

Too many characters.

Enter the correct number of characters.

8

Incorrect input mode.

See the Communication Interface User’s Manual.

9

Incorrect input range code.

See the Communication Interface User’s Manual.

10

A disabled loop number is selected.

See the Communication Interface User’s Manual.

11

A disabled module number is selected.

See the Communication Interface User’s Manual.

12

A disabled input kind is selected.

See the Communication Interface User’s Manual.

13

A disabled pid number is selected.

See the Communication Interface User’s Manual.

14

A disabled segment number is selected.

See the Communication Interface User’s Manual.

21

Cannot set an alarm for a skipped channel.

See the Communication Interface User’s Manual.

22

The upper and lower span limits are equal.

See sections 4.6 and 7.1.

23

The upper and lower scale limits are equal.

See sections 4.6 and 7.1.

30

The partial boundary value exceeds the range of the span.

See section 8.14.

31

Partial-expansion display is set ON for a SKIPPED channel.

See the Communication Interface User’s Manual.

35

The upper and lower limits of the display band are equal.

See section 8.12.

36

The lower limit of the display band is greater than the upper limit.

See section 8.12.

37

The display band is narrower than 4% of the entire display.

See section 8.12.

40

Incorrect group set character string.

See sections 6.1 and 8.3.

41

There is no specified input channel.

See the Communication Interface User’s Manual.

42

Exceeded the number of channels which can be set.

See the Communication Interface User’s Manual.

43

A channel number cannot repeat in a group.

See section 8.3.

45

There is no character string saved in the clipboard.

Copy a character string to the clipboard.

46

The character string saved in the clipboard is too long.

Paste a character string with the specified number
of characters.

61

There is no channel specified by the MATH expression.

See section 10.1.

62

MATH expression grammar is incorrect.

See section 10.1.

63

MATH expression sequence is incorrect.

See section 10.1.

64

MATH upper and lower span values are equal.

See section 10.1.

70

The range of the MATH constant is exceeded.

See section 10.1.

71

Set range of the MATH constant is exceeded.

See section 10.1.

81

All space or ‘quit’ string cannot be specified.

See section 11.3.

85

The login password is incorrect.

See section 11.3.

86

The key-lock release password is incorrect.

See section 11.2.

87

This key is locked.

See section 11.2.

88

This function is locked.

See section 11.2.

89

Press [FUNC] key to login.

See section 11.3.

90

No permission to enter to the SETUP mode.

See section 11.3.

91

Password is incorrect.

See sections 11.2 and 11.3.