1.14 Measurement Function > Measurement Input

Related Settings

Integration Time of the A/D Converter

The CX2000 uses an A/D converter to convert the sampled analog signal to a digital
signal. At this point, the sampled data is integrated for a certain period to eliminate the
noise that is mixed in the input signal. You can select the integral time from [Auto]/
[50Hz(20ms)]/[60Hz(16.7ms)/100ms]. This setting applies not only to the measurement
input of the measurement function but also to the PV input of the control function. By
setting the integration time of the A/D converter to match the time period corresponding
to one cycle of the power supply or an integer multiple of one cycle, the power supply
frequency noise can be effectively eliminated. If [Auto] is selected, the recorder will
automatically detect the power supply frequency and select 16.7 ms or 20 ms. Because 100 ms
is an integer multiple of 16.7 ms and 20 ms, this setting can be used to eliminate the power
frequency noise for either frequency, 50 Hz or 60 Hz.

Scan Interval

The scan interval is the interval used to sample the input signal. You can select [1s] or
[2s]. However, if the integral time is set to [100ms], the scan interval is fixed to 2 s.

Note

When performing four arithmetic operations on models with the computation function (/M1),

this scan interval is used to carry out the computation.

Input Type and Input Computation (Mode)

If the last two digits of the CX2000 model that you are using are 1 and 0, there are 10 channels of
measurement input channels (measurement channels); if they are 2 and 0, there are 20
channels. You can select the type of signal input to each measurement channel from DC
voltage, thermocouple, resistance temperature detector, and ON/OFF input (contact signal or
voltage signal). In addition, difference computation, square-root computation, and scaling can be
performed on the measured data and display or save the computed result as measured data.
On the CX2000, the input type and input computation type is set as a [Mode]. In
addition, if [Mode] is set to difference computation, square-root computation, or scaling,
the input type is set as a [Type].

Mode

Notation in Setup*

Description

DC voltage

Voltage

Measures a DC voltage in the range of ±20 mV to ±50 V.

Thermocouple

TC

Measures temperatures corresponding to the temperature range of each
thermocouple type such as R, S, B, K, E, J, T, N, and W.

Resistance

RTD

Measures temperatures corresponding to the

temperature detector

appropriate range for Pt100 or JPt100.

Scaling

Scale

The input signal can be scaled to a value in the appropriate unit and displayed. You
can select the input type from DC voltage, thermocouple, resistance temperature
detector, and ON/OFF input.

Difference

Delta

Displays the value obtained by subtracting the measured value of another channel
(called the reference channel) from the input signal of the specified channel as the
measured value for the specified channel. You can select the input type from DC
voltage, thermocouple, resistance temperature detector, and ON/OFF input.

ON/OFF input

DI

Displays the contact input or voltage input signals by correlating them to 0% or
100% of the display range.
Contact input: Closed contact is ON (1).

Open contact is OFF (0).

Voltage input: Less than 2.4 V is OFF (0).

Greater than or equal to 2.4 V is ON (1)

Square root

Sqrt

Calculates the square root of the input signal and

computation

displays the result as the measured value of the channel. The computed
result can also be scaled to a value in the appropriate unit and displayed.
The input type is DC voltage only.

Skip

Skip

Disables the channel.