• When set to [TC] (thermocouple input)

Range

Selectable Measurement Span Range

R

0.0 to 1760.0 °C

S

0.0 to 1760.0 °C

B

0.0 to 1820.0 °C

K

–200.0 to 1370.0 °C

E

–200.0 to 800.0 °C

J

–200.0 to 1100.0 °C

T

–200.0 to 400.0 °C

N

0.0 to 1300.0 °C

W

0.0 to 2315.0 °C

L

–200.0 to 900.0 °C

U

–200.0 to 400.0 °C

PLATI

0.0 to 1400.0 °C

PR

0.0 to 1900.0 °C

WRe

0.0 to 2400.0 °C

• When set to RTD (resistance temperature detector)

Range

Selectable Measurement Span Range

Pt100

–200.0 to 600.0 °C

JPt100

–200.0 to 550.0 °C

• When set to [DI] (ON/OFF input)

Range

Selectable Measurement Span Values

Level

0: Less than 2.4 V,
1: Greater than or equal to 2.4 V

Cont

0: Open, 1: Closed

• When set to [Delta] (difference computation)

Type

Range

Selectable Measurement Span Range

Voltage

20mV

–20.00 to 20.00 mV

60mV

–60.00 to 60.00 mV

200mV

–200.0 to 200.0 mV

2V

–2.000 to 2.000 V

6V

–6.000 to 6.000 V

20V

–20.00 to 20.00 V

50V

–50.00 to 50.00 V

TC

R

–1760.0 to 1760.0 °C

S

–1760.0 to 1760.0 °C

B

–1820.0 to 1820.0 °C

K

–1570.0 to 1570.0 °C

E

–1000.0 to 1000.0 °C

J

–1300.0 to 1300.0 °C

T

–600.0 to 600.0 °C

N

–1300.0 to 1300.0 °C

W

–2315.0 to 2315.0 °C

L

–1100.0 to 1100.0 °C

U

–600.0 to 600.0 °C

PLATI

–1400.0 to 1400.0 °C

PR

–1900.0 to 1900.0 °C

WRe

–2400.0 to 2400.0 °C

RTD

Pt100

–800.0 to 800.0 °C

JPt100

–750.0 to 750.0 °C

DI

Level

–1 to 1

Cont

–1 to 1

7.1 Settings Related to Measurement Inputs