9-7

IM 04L31A01-01E

Data Save/Load Operations

9

9.2

Saving Measured/Computed Data at Arbitrary
Times (Manual Sample)

This section explains how to save the instantaneous values of all measurement/
computation channels (excluding measurement channels set to skip and computation
channels set to Off) at arbitrary times.

Procedure

1. On the operation dislay, press the

FUNC key

.

A soft key menu is displayed at the bottom of the display.

2. Press the

[Manual sample] soft key

.

Note

•

When manual sample is executed, the instantaneous values of all channels (excluding

measurement channels set to skip and computation channels set to Off) are acquired to

the internal memory.

•

The first time manual sample is executed, a manual sampled data file is created on the

external storage medium. The data is appended to this file for each successive manual

sample operation.

•

Up to 50 sets of data can be acquired to the internal memory. When 50 is exceeded, the

data is overwritten from the oldest data.

•

The number of manual sample data points in the internal memory can be confirmed using

memory summary (see

section 1.16

).

•

For the data format of the manual sampled data, see

appendix 5, “Data Formats of ASCII

Files.”

•

You can assign the manual sample function to the USER key and save the data simply by

pressing a single key. For the procedure in assigning the function to the USER key, see

section 11.1

.