11-13

IM 04L31A01-01E

Operations of Other Functions

11

Manual sample
• Remote input signal: Trigger, 250 ms or more.
• Write the instantaneous values of all channels to the internal memory once. This

excludes measurement channels set to skip and computation channels set to Off.

Panel1Load to Panel3Load
• Remote input signal: Trigger, 250 ms or more.
• Loads the setup data file named “LOAD1.PCL”, “LOAD2.PCL”, or “LOAD3.PCL”

stored on the external storage medium and activates the settings. Setup data files
named “LOAD1.PCL”, “LOAD2.PCL”, or “LOAD3.PCL” must be created and saved
to the external storage medium beforehand.

Message1 to Message8
• Remote input signal: Trigger, 250 ms or more.
• Displays a message at the position corresponding to the time when the signal was

applied on the trend display. The displayed message is also written to the internal
memory.
When data acquisition to the internal memory is stopped, messages cannot be
displayed or written. Applying a remote input signal produces no effect.

Snapshot
• Remote input signal: Trigger, 250 ms or more.
• Saves the current screen image data to the external storage medium. The

snapshot function operates in all modes (operation mode, setting mode, and basic
setting mode). Error messages, even if they are displayed, are not saved.

11.5 Setting the Measurement Remote Input (/A6R option or /A4FR option)