Standard performance – Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 367

Advertising
background image

14-19

IM 04L31A01-01E

Specifications

14

Isolation
Insulation resistance:

Between each terminal and earth: 20 MΩ or more (at 500 VDC)

Withstand voltage:

Between the power terminal and earth:

1500 VAC at 50/60 Hz for one minute

Between the contact output terminal and earth:

1500 VAC at 50/60 Hz for one minute

Between the measurement input terminal and earth:

1500 VAC at 50/60 Hz for one minute

Mutually between measurement input terminals:

1000 VAC at 50/60 Hz for one minute

Between the contact input terminal and earth:

500 VDC for one minute

Between the current output terminal and earth:

500 VAC at 50/60 Hz for one minute

Between the voltage pulse output terminal and earth:

500 VAC at 50/60 Hz for one minute

Between the transistor contact output and earth:

500 VDC for one minute

Grounding resistance:

100 Ω or less

Standard Performance

Measurement accuracy: The following specifications apply to operation of the CX2000

under standard operation conditions:
Temperature:

23 ± 2 °C

Humidity:

55% ± 10% RH

Supply voltage: 90 to 132 or 180 to 250 VAC
Power frequency: 50/60 Hz ± 1% Warm-up time: At least 30 minutes.
Other ambient conditions such as vibration should not
adversely affect the operation of the CX2000.

Input Type

DCV

TC
(Excluding the
reference
junction
compensation
accuracy)

RTD

Measurement Accuracy (Digital Display)

±(0.1% of rdg + 2digits)

±(0.1% of rdg + 3digits)
±(0.15% of rdg + 1°C)
Except R, S: 0 to 100°C: ±3.7°C,
100.0 to 300°C: ±1.5B: 400 to 600°C: ±2°C,
Accuracy at less than 400°C is not warranted.
±(0.15% of rdg + 0.7°C)
Except ±(0.15% of rdg + 1°C) at –200 to –100°C
±(0.15% of rdg + 0.5°C)
±(0.15% of rdg + 0.5°C)
Except ±(0.15% of rdg + 0.7°C) at –200 to –100°C
±(0.15% of rdg + 0.7°C)

±(0.15% of rdg + 1°C)
±(0.15%of rdg + 0.5°C)
Except ±(0.15% of rdg + 0.7°C at –200 to 100°C)
±(0.25% of rdg ± 2.3°C)
0 to 450°C Accuracy

not

guanteed

450 to 750°C ±(0.9% of rdg + 16.0°C)
750 to 1100°C ±(0.9% of rdg + 6.0°C)
1100 to 1900°C ±(0.9% of rdg + 2.0°C)
±(0.3% of rdg ± 2.8°C)

±(0.15% of rdg + 0.3°C)

Max.Resolution of Digital Display

10 µV
10 µV

100 µV

1mV

10mV
10mV

0.1°C

Range
20 mV
60 mV

200 mV

2 V
6 V

20 V
50 V

R

S

B

K

E

J

T

N

W

L

U

PLATINEL

PR40-20

W3Re/W25Re

Pt100

JPt100

1mV

Measurement accuracy during scaling:

Accuracy during scaling (digits) = measurement accuracy
(digits) × multiplier + 2 digits (rounded up)
* Fractions rounded up

Reference junction compensation:

INT (internal) and EXT (external) switchable (common to all
channels)

14.8 General Specifications