Standard Installation Environment

Standard operating conditions:

Ambient temperature: 0 to 50 °C (5 to 40 ° C when floppy disk or Zip disk is in operation)
Ambient humidity:

20 to 80% RH (at 5 to 40 °C)

Vibration:

10 to 60 Hz 0.2 m/s

2

Shock:

Not allowed

Magnetic field:

400 A/m or less (DC and 50/60 Hz)

External noise:

Normal mode (50/60 Hz)

Volt: The peak value including the signal must be less than

1.2 times the measuring range.

TC: The peak value including the signal must be less than

1.2 times the measuring thermal electromotive force.

RTD: 50 mV or less

Common mode noise (50/60 Hz):

250 Vrms AC or less for all ranges

Maximum noise voltage between channels (50/60 Hz):

250 Vrms AC or less

Mounting position:

Up to 30° of backward tilt angle allowed, no tilt to the left or right

Warm-up time:

At least 30 minutes after power-up

Altitude:

2000 m or less

Transport and storage conditions:

Ambient temperature: –25 to 60 °C
Humidity:

5 to 95%RH (no condensation)

Vibration:

10 to 60 Hz 4.9 m/s

2

Shock:

392 m/s

2

or less (packaged condition)

Safety and EMC Standards
Safety

CSA:

Certified by CSA22.2 No. 61010.1, Installation category II

*1

,

Pollution degree 2

*2

, Measurement category II

*3

UL:

Certified by UL61010-1 (CSA NRTL/C)

CE:

Complies with EN61010-1

*

1 “Installation category (Overvoltage category)” describes a number which defines a transient

overvoltage condition. It implies the regulation for impulse withstand voltage. “II” applies to
electrical equipment which is supplied from the fixed installation like distribution board.

*

2 “Pollution degree” describes the degree to which a solid, liquid, or gas which deteriorates

dielectric strength or surface resistivity is adhering. “2” applies to normal indoor atmosphere.
Normally, only non-conductive pollution occurs.

*

3 Applies to measuring circuits connected to low voltage installation, and electrical instruments

supplied with power from fixed equipment such as electric switchboards.

EMC Conformity standards

Altitude:

2000 m or less

CE:

EN61326-1 Class A, Table 2 (For use in industrial locations)
EN61326-2-3
EN55011 Class A, Group1
EN61000-3-2 Class A
EN61000-3-3

C-tick:

EN55011 Class A, Group1

The instrument continues to operate at a measurement accuracy of within ±10% of
the range during testing.

Power Supply
Supply voltage:

100 to 110 VAC±10%, 200 to 220 VAC±10%

Power frequency:

50 Hz ± 2%, 60 Hz ± 2%

Power consumption:

Supply Voltage

100 VAC
240 VAC

LCD Saver Mode

Approx. 43 VA
Approx. 62 VA

Normal

Approx. 45 VA
Approx. 65 VA

Max.

75 VA

106 VA

* For the power supply specifications when the CX1000 is driven by the 24-V power

supply, see “24 VDC/AC power supply driven (/P1)” in section 14.7, “Options.”

14.8 General Specifications