5 pre/post screen deposit page parameters – INFICON IC6 Thin Film Deposition Controller User Manual

Page 120

Advertising
background image

5 - 12

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

NOTE: After a crystal fails due to exceeding the Counts value, the

Crystal Fail can be cleared by installing a new crystal or
pressing the F3 key to Clear S & Q counts on the Sensor
Information screen. This action must be repeated for each
sensor used by cursoring to the appropriate sensor number.

STABILITY SINGLE . . . . . . 0 Hz and 25 to 9999 Hz (Values 1 to 24 are

not allowed)

This sets the maximum permissible increase in frequency from one
measurement to the next that will cause a crystal fail. The default value
is 0 Hz which disables the function.

STABILITY TOTAL . . . . . . . 0 Hz and 25 to 9999 Hz (Values 1 to 24 are

not allowed)

This sets the maximum total permissible increase in frequency
accumulated during the active layer that will cause a crystal fail. The
default value is 0 Hz which disables the function.

NOTE: After a crystal fails due to exceeding the allowable Hz value,

the Crystal Fail can be cleared by installing a new crystal or
pressing the F3 key to Clear S & Q counts on the Sensor
Information screen. This action must be repeated for each
sensor used by cursoring to the appropriate sensor number.

5.1.5 Pre/Post Screen Deposit Page Parameters

MATERIAL NUMBER . . . . . . . . . . . 1 to 32

Change the number to edit or create a different material

NOTE: The first power ramp consisting of Pre-Cond Rise and Pre-Cond Soak is

skipped if the previous layer had the same source, same crucible and its
Idle power is nonzero.

PRECON SOAK POWER . . . . . . . . 0.00 to 99.99%

This parameter is usually set to the power level at which the source material
just begins to glow. The IC6 ramps the power level from zero to Pre-Cond Soak
Power linearly over the time period Pre-Cond Rise Time. The default value is 0.

PRECON RISE TIME. . . . . . . . . . . . 00:00 to 99:59 min:s

This parameter provides the time period over which the source power is
ramped from 0 to Pre-Cond Soak Power. The default values is 00:00.

PRECON SOAK TIME. . . . . . . . . . . 00:00 to 99:59 min:s

This parameter provides the time period for which the IC6 holds at
Precon Soak Power. The default value is 00:00.