6 query general parameter, 7 update general parameter – INFICON IC6 Thin Film Deposition Controller User Manual

Page 184

Advertising
background image

10 - 14

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

10.4.6 Query General Parameter

Command = QG <Command ID>
Command ID = <Byte> See Cmnd ID in

Table 10-6

Response = <Integer> | <Float>
Description = Type of response depends on command ID, see Data Type column
in

Table 10-6

.

10.4.7 Update General Parameter

Command = UG <Command ID> <Parameter Value>

Command ID = <Byte> See

Table 10-6

Parameter Value = <Integer> | <Float> Type of value depends on command ID,
see Data Type column in

Table 10-6

.

Response = None (Just header and trailer)

General Parameters

:

Table 10-6 General parameters

QG UG

Cmd ID

Name

(Units)

Range:

Low

Limit/

High

Limit/

Default

Data Type

Notes

Update Restrictions

and Display Notes

1

(0x01)

Active Process

1/ 50/ 1

Integer

Update not allowed if

process is running.

2

(0x02)

Layer To Start

1/ 200/ 1

Integer

Cannot be greater than

number of layers in active

process.

If process change causes this

conflict, then value will be

changed to 1.

3

(0x03)

Run Number

1/ 9999/ 1

Integer

4

(0x04)

Auto Start Next

0/ 1/ 0

Encode

0 = No

1 = Yes

5

(0x05)

Max Concurrent

Layers

1/ 6/ 2

Integer

Number of layers allowed

running simultaneously

6

(0x06)

Layers Displayed

0/ 2/ 0

Encode

0 = Auto

1 = 4 Layers
2 = 6 Layers

Gray Font if Max Concurrent

Layer is < 3

7

(0x07)

Not Used

Advertising