Ble 10-11, Table 10-11 – INFICON IC6 Thin Film Deposition Controller User Manual

Page 208

Advertising
background image

10 - 38

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

Table 10-11 Sensor parameters

QS US

Cmd ID

Name

(Units)

Range:

Low Limit/

High Limit/

Default

Data Type

Notes

Update Restrictions

and Display Notes

1

(0x01)

(Sensor) Shutter

Out(put)

0/ 38/ 0

Integer

0 = not defined

Update not allowed if the sensor is running. Not accepted

if output in use by other hardware

2

(0x02)

Sensor Type

0/ 4/ 0

Encode

0 = Single 1 = XtalTwo

2 = XtalSix 3 = Xtal12

4 = Generic

Update not allowed if the

sensor is running.

3

(0x03)

(Sensor) Switch

Out(put)

0/ 38/ 0

Integer

0 = not defined

Gray font if Sensor Type =

Single

Update not allowed if the

sensor is running. Not

accepted if output in use

by other hardware.

4

(0x04)

Auto-Z

0/ 1/ 0

Encode

0 = No, 1 = Yes

5

(0x05)

# of Positions

1/ 12/ 1

Integer

Visible only when Sensor

Type = 4

Update not allowed if the

sensor is running.

6

(0x06)

# of Pulses

1/ 10/ 1

Integer

7

(0x07)

Pulse On (Time)

(Seconds)

0.1/ 9.9 /1

REAL_3X1

8

(0x08)

Pulse Off (Time)

(Seconds)

0.1/ 9.9 /1

REAL_3X1

9

(0x09)

Recorder Output

0/ 12/ 0

Integer

0 = not defined

Not allowed to be the

same as a source DAC

10

(0x0a)

Recorder Function

0/ 2/ 0

Encode

0 = Rate 1 = Thick

2 = RateDev

Note: Rate Filter Time has

no effect

11

(0x0b)

Recorder Range

0/

99999/

100

Integer

Gray font when Recorder

Function = Rate Deviation