Section 15.9.1.3 on – INFICON IC6 Thin Film Deposition Controller User Manual

Page 304

Advertising
background image

15 - 34

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

12

Verify electrical continuity from the center pin of the Microdot connector on the
feedthrough to the un-terminated end of the six-inch BNC cable.

13

Verify electrical isolation from the center pin to electrical ground (feedthrough
body).

If the feedthrough/six-inch BNC cable system is found to be defective, look for
defective contacts at the feedthrough to BNC cable connection. Repair or replace
the feedthrough as necessary, re-attach the BNC cable to the XIU and in-vacuum
cable to the Crystal head and repeat this procedure starting at step 2.

15.9.1.3 Sensor Head Or Monitor Crystal

Diagnostic Procedure

NOTE: The procedure is for use with front load style sensor heads.

1

Remove the Crystal Cover from the Sensor Head.

2

Refer to

Figure 15-7 on page 15-31

. Connect the Crystal Sensor Emulator to

the Sensor Head at Point C.

If the Crystal Fail message disappears after approximately 5 seconds the
Sensor Head is operating properly. Remove the Crystal Sensor Emulator
and re-insert the Crystal Cover into the Sensor Head.

If the Crystal Fail message remains, continue at step 3.

3

Disconnect the in-vacuum cable from the Sensor Head and the feedthrough.
Remove the Crystal Sensor Emulator from the Sensor Head.

4

Using an Ohm meter, verify the electrical connections on the Sensor Head.

Verify there is electrical continuity from the center pin contact on the
Microdot connector on the Sensor Head to the leaf spring contact in the
Sensor Head. Take care not to apply too much pressure on the center pin
of the microdot connector as it may become damaged.

There must be electrical isolation between the center pin of the Microdot
connector and the Sensor Head body.

If the Sensor Head is found to be defective, contact INFICON to have the
Sensor Head repaired.

Advertising