16 query material parameter, 17 update material parameter – INFICON IC6 Thin Film Deposition Controller User Manual
Page 196
10 - 26
PN
07
4-
50
5-
P1
F
IC6 Operating Manual
10.4.16 Query Material Parameter
Command = QM <Command ID> <Material Number>
Command ID = <Byte> See
Material Number = <Byte> 1- 32
Response = <Integer> | <Float>
Type of response depends on command ID, see Data Type column in
.
10.4.17 Update Material Parameter
Command = UM <Command ID> <Material Number> <Parameter Value>
Command ID = <Byte> See
Material Number = <Byte> 1- 32
Parameter Value = <Integer> | <Float> Type of value depends on command ID,
see Data Type column in
.
Response = None (Just header and trailer)
Table 10-9 Material parameters
QM UM
Cmd ID
Name
(Units)
Range:
Low Limit/
High Limit/
Default
Data Type
Notes
Update Restrictions
and Display Notes
1
(0x01)
Density
(g/cc)
0.1/ 99.999/
1.0
REAL_5X3
2
(0x02)
Z-Ratio
0.1/ 15.0/
1.0
3
(0x03)
Master Tooling
(%)
1.0/ 999.9/
100.0
REAL_4X1
4
(0x04)
Control Loop
0/ 2/ 0
Encode
0 = NonPID 1 = PI
2 = PID
5
(0x05)
Process Gain
(Å/s/%)
0.01/
999.99/ 10.0
REAL_
5F32
6
(0x06)
Time Constant
(seconds)
0.01/
9999.99/ 1.0
REAL_
6F32
Gray font if Control Loop =
NonPID
7
(0x07)
Dead Time
(seconds)
0.01/
9999.99/ 1.0
REAL_
6F32
Gray font if Control Loop =
NonPID
8
(0x08)
Source
1/ 6/ 1
Integer
Update not allowed if
material is running.
9
(0x09)
Maximum Power
(%)
0.01/ 99.99/
90.0
REAL_4X2