16 query material parameter, 17 update material parameter – INFICON IC6 Thin Film Deposition Controller User Manual

Page 196

Advertising
background image

10 - 26

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

10.4.16 Query Material Parameter

Command = QM <Command ID> <Material Number>

Command ID = <Byte> See

Table 10-9

Material Number = <Byte> 1- 32

Response = <Integer> | <Float>

Type of response depends on command ID, see Data Type column in

Table 10-9

.

10.4.17 Update Material Parameter

Command = UM <Command ID> <Material Number> <Parameter Value>

Command ID = <Byte> See

Table 10-9

.

Material Number = <Byte> 1- 32

Parameter Value = <Integer> | <Float> Type of value depends on command ID,
see Data Type column in

Table 10-9

.

Response = None (Just header and trailer)

Table 10-9 Material parameters

QM UM
Cmd ID

Name

(Units)

Range:

Low Limit/

High Limit/

Default

Data Type

Notes

Update Restrictions

and Display Notes

1

(0x01)

Density

(g/cc)

0.1/ 99.999/

1.0

REAL_5X3

2

(0x02)

Z-Ratio

0.1/ 15.0/

1.0

3

(0x03)

Master Tooling

(%)

1.0/ 999.9/

100.0

REAL_4X1

4

(0x04)

Control Loop

0/ 2/ 0

Encode

0 = NonPID 1 = PI

2 = PID

5

(0x05)

Process Gain

(Å/s/%)

0.01/

999.99/ 10.0

REAL_

5F32

6

(0x06)

Time Constant

(seconds)

0.01/

9999.99/ 1.0

REAL_

6F32

Gray font if Control Loop =

NonPID

7

(0x07)

Dead Time

(seconds)

0.01/

9999.99/ 1.0

REAL_

6F32

Gray font if Control Loop =

NonPID

8

(0x08)

Source

1/ 6/ 1

Integer

Update not allowed if

material is running.

9

(0x09)

Maximum Power

(%)

0.01/ 99.99/

90.0

REAL_4X2

Advertising