24 query sensor parameter, 25 update sensor parameter – INFICON IC6 Thin Film Deposition Controller User Manual

Page 207

Advertising
background image

10 - 37

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

10.4.24 Query Sensor Parameter

Command = QS <Command ID> <Sensor Number>

Command ID = <Byte> See

Table 10-11

.

Sensor Number = <Byte> Sensor number 1 - 8

Response = <Integer> | <Float>

Type of response depends on command ID, see Data Type column in

Table 10-11

.

10.4.25 Update Sensor Parameter

Command: = US <Command ID> <Sensor Number> <Parameter Value>

Command ID = <Byte> See

Table 10-11

Sensor Number = <Byte> Sensor number 1 - 8

Parameter Value = <Integer> | <Float> Type of value depends on command ID,
see Data Type column in

Table 10-11

.

Response = None (Just header and trailer)

4

(0x04)

(Crucible)

Cruc

1/ 64/ 1

Integer

Update not allowed if active

process is running

5

(0x05)

(Co-Deposition)

CoDep

0/ 1/ 0

Encode

0 = No 1 = Yes

When this layer is set to yes

the next layer will not

display.Gray font if next layer

is empty OR next layer is set

to Yes OR this layer is the last

layer.

Update not allowed if active

process is running No update

if the last programmed layer

OR if the next layer's Co-Dep

parameter is set to Yes

6

(0x06)

(Ratio Control)

Ratio Cntrl

(%)

0/

999.9/

0

REAL_4X1

Gray Font when

Co-deposition is No

Table 10-10 Process layer parameters (continued)

QP UP

Cmd ID

Name

(Units)

Range:

Low Limit/

High Limit/

Default

Data Type

Notes

Update Restrictions

and Display Notes

Advertising