IC6 Operating Manual

PROTOCOL . . . . . . . . . . . . . . . . . . . Standard (0), Dlog Page (1), Dlog Comma (2)

Standard enables receiving communication commands from an external host
and sending the expected responses. Dlog Comma and Dlog Page allow only
one-way transmission of information from the IC6 to the external device. The
Dlog Comma format allows importation of the Data Log string into a
spreadsheet program. Dlog Page format is designed for use as text only. A
carriage return (d13, h0D) and line feed (d10, h0A) are inserted at the end of
each line of data. The Dlog information is transmitted when the source shutter
closes at the end of Deposit or following a Stop or a Suspend while in Deposit.
The default is Standard (0).

If the Ethernet option is installed, the following parameters will be shown

IP ADDRESS . . . . . . . . . . . . . . . . . . nnn.nnn.nnn.nnn

Use the up and down cursor keys to scroll through the address and the
numerics keys to enter the desired address. The default address is
10.211.72.203.

NET MASK. . . . . . . . . . . . . . . . . . . . nnn.nnn.nnn.nnn

The net mask is some number of leading bits set to 1, followed by all 0. The
default net mask value is 255.255.0.0:

255

.255

. 0

.0

11111111 1111111 00000000 00000000