8 query input name, 9 update input name, 10 query process name – INFICON IC6 Thin Film Deposition Controller User Manual
Page 188: 11 update process name
10 - 18
PN
07
4-
50
5-
P1
F
IC6 Operating Manual
10.4.8 Query Input Name
Command = QI <Input Number>
Input Number = <Byte> 1 - 28
Response = <Input Name>
Input Name = <String> Up to 15 characters, null terminated
10.4.9 Update Input Name
Command = UI <Input Number> <Input Name>
Input Number = <Byte> 1 - 28
Input Name = <String> Up to 15 characters, null terminated
Response = None (Just header and trailer)
Input Names may only be updated if the IC6 is in Ready or Stop.
NOTE: UI changes are not allowed while a process is running. The UI command
will not allow updating a name set by hardware designation, unless it is
identical to the defined hardware name.
10.4.10 Query Process Name
Command = QK < Process Number>
Process Number = <Byte> 1 - 50
Response = < Process Name>
Process Name = <String> Up to 15 characters, null terminated. The string
returned will be a blank if the default Process # name has not been changed.
10.4.11 Update Process Name
Command = UK <Process Number> <Process Name>
Process Number = <Byte> 1 - 50
Process Name = <String> Up to 15 characters, null terminated
Response = None (Just header and trailer)