10.4.18 Query Material Name
Command = QN <Material Number>
Material Number = <Byte> 1-32
Response = < Material Name>
Material Name = <String> Up to 15 characters, null terminated. The string returned
will be a blank if the default Material # name has not been changed.
10.4.19 Update Material Name
Command = UN < Material Number> < Material Name>
Material Number = <Byte> 1 - 32
Material Name = <String> Up to 15 characters, null terminated
Response = None (Just header and trailer)
10.4.20 Query Output Name
Command = QO <Output Number>
Output Number = <Byte> 1 - 38
Response = <Output Name>
Output Name = <String> Up to 15 characters, null terminated
10.4.21 Update Output Name
Command = UO <Output Number> <Output Name>
Output Number = <Byte> 1 - 38
Output Name = <String> Up to 15 characters, null terminated
Response = None (Just header and trailer)
Output Names may only be updated if the IC6 is in Ready or Stop.
NOTE: UO changes are not allowed while a process is running. The UO command
will not allow updating a name set by hardware designation, unless it is
identical to the defined hardware name