19 update input name, 20 query output name, 21 update output name – INFICON IC6 Thin Film Deposition Controller User Manual
Page 227: 22 query output type, 23 update output type

10 - 57
PN
07
4-
50
5-
P1
F
IC6 Operating Manual
10.4.35.19 Update Input Name
Input 1, “P1”
Command Format: UI <Input Number> <Input Name>
UI 1 P1
Command: 060055490150310020
Response: 030000A306A9
10.4.35.20 Query Output Name
Output 1
Command Format: QO <Output Number>
QO 1
Command Q0 1: 0300514F01A1
Response: 0B0000E206534855545445520017
10.4.35.21 Update Output Name
Output 1, “SHUTTER”
Command Format: UO <Output Number> <Output Name>
UO 1 SHUTTER
Command U0 1: 0B00554F015348555454455200D4
Response: 030000E306E9
10.4.35.22 Query Output Type
Output 1
Command Format: QT <Output Number>
QT 1
Command QT 1: 0300515401A6
Response: 040000BF0600C5
10.4.35.23 Update Output Type
Output 1, Normally Closed (1)
Command Format: UT <Output Number> <Output Type>
UT 1 1
Command UT 1 1: 040055540101AB
Response: 030000FA0600