2 cool drawer – INFICON IC6 Thin Film Deposition Controller User Manual
Page 297

15 - 27
PN
07
4-
50
5-
P1
F
IC6 Operating Manual
15.8.2 Cool Drawer
Follow the procedure below to replace the crystal in a Next Generation Cool
Drawer™ sensor:
1
Using your thumb and index fingers, gently squeeze the sides of the retainer at
the mid section then lift it up, away from the drawer, as shown in
.
2
Hold the drawer by the handle and turn it upside down to remove the spent
crystal.
3
Install a new crystal in the drawer. Observe its orientation. The pattern
electrode should face upward as shown in
4
Hold the retainer by its sides. Align its orientation notch with the drawer then
gently and evenly push the retainer down until it snaps firmly into the drawer.
see
. Never push down (or pull up) on the contact spring, doing so
may permanently damage it.
5
Inspect the whole assembly. The retainer should be even and engage the
drawer at all four corners.
Figure 15-2 Cool drawer - removing the crystal
Figure 15-3 Cool drawer - replacing the crystal
Contact Spring
Orientation
Notch
Handle
Retainer
Crystal
Drawer