INFICON IC6 Thin Film Deposition Controller User Manual

Page 199

Advertising
background image

10 - 29

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

38

(0x26)

On Final Thickness

0/ 2/ 1

Encode

0 = Continue, 1 = Post-Dep

2 = NonDepCont

39

(0x27)

Ramp 1 Rate

(Å/s)

0/ 999.9/ 0

REAL_

4F31

40

(0x28)

Start Ramp 1

(KÅ)

0/ 999.999/

0

REAL_

4F31

Value must be less than Start Ramp 2 thickness value if

Start Ramp 2 thickness value is non-zero.

41

(0x29)

Ramp 1 Time

(seconds)

(mm:ss)

0/ 5999/ 0

0/ 99:59/ 0

Time

See Time Note

42

(0x2a)

Ramp 2 Rate

(Å/s)

0/ 999.9/ 0

REAL_

4F31

43

(0x2b)

Start Ramp 2

(KÅ)

0/ 999.999/

0

REAL_

4F31

Value must be greater than Start Ramp 1 thickness value.

44

(0x2c)

Ramp 2 Time

(seconds)

(mm:ss)

0/ 5999/ 0

0/ 99:59/ 0

Time

See Time Note

45

(0x2d)

RateWatcher Option

0/ 1/ 0

Encode

0 = No, 1 = Yes

46

(0x2e)

RateWatcher Time

(seconds)

(mm:ss)

0/ 5999/ 0

0/ 99:59/ 0

Time

See Time Note

Grey font unless

RateWatcher option set to

Yes.

47

(0x2f)

RateWatcher

Accuracy (%)

1/ 99/ 5

Integer

48

(0x30)

Multipoint

0/ 1/ 0

Encode

0 = No, 1 = Yes

Update not allowed if

material is running.

49

(0x31)

Aggregate Multiplier

0/ 1/ 0

Encode

0 = No, 1 = Yes

Visible only if Multipoint is

set to Yes

Table 10-9 Material parameters (continued)

QM UM
Cmd ID

Name

(Units)

Range:

Low Limit/

High Limit/

Default

Data Type

Notes

Update Restrictions

and Display Notes

Advertising