4 cross talk calibration select display – INFICON IC6 Thin Film Deposition Controller User Manual

Page 261

Advertising
background image

12 - 19

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

12.6.4 Cross Talk Calibration SELECT display

After the RAMP down state the IC6 enters the SELECT display, from which the
user can accept or reject the calibration thickness values, as desired. If the values
are accepted they are stored on the Material’s Sensor page.

Figure 12-11 Crosstalk complete display

12.6.4.1 Function Key Selection Choices for

Cross Talk Calibration Select

Key Function

Description

F3

ACCEPT

Press this key to accept the Calibration Thickness
values and returns to the Cross Talk Calibration
SETUP display. The accepted Calibration Thickness
values are entered into page 3 of the Material Set-Up
display.

F4

REJECT

Press this key to reject the Calibration Thickness
values and return to the Cross Talk Calibration
SETUP display.