10.4.33 Remote General Action

Command = RG <Command ID> (<Action Value>)

Command ID = <Byte> See

Table 10-15

.

Action Value = <Byte>… <Byte> | <Integer>…<Integer>|<Float>…<Float>

See

Table 10-15

for commands that require an action value.

Response = None (Just header and trailer)

Table 10-15 Remote general

RG

Command

ID

Meaning

Action Value

Notes

0 (0x00)

Start

Duplicates the equivalent front panel action

1 (0x01)

Stop

Duplicates the equivalent front panel action

2 (0x02)

Reset

Duplicates the equivalent front panel action

3 (0x03)

Remote Lock On

No parameters can be updated from the front panel

"R Lock" displayed

4 (0x04)

Remote Lock Off

Cancel RG 3 command

5

(0x05)

Set Logic Statement

<Statement Number>

to True

<Byte>

Range is 1-100.

Only does something if logic statement event "Computer

Control" set. Otherwise, ignored.

6

(0x06)

Clear Logic

Statement

<Statement Number>

to False.

<Byte>

Range is 1-100.

Only does something if logic statement event "Computer

Control" set. Otherwise, ignored.

7

(0x07)

Crystal Fail Inhibit On

<Byte>

Range is 0-8 (0 = All)

"Crystal Fail" output always set to FALSE, even during crystal

fail

8

(0x08)

Crystal Fail Inhibit Off

<Byte>

Range is 0-8 (0 = All)

"Crystal Fail" output has normal meaning

9

(0x09)

Set date (Month, Day,
Year) or (Day, Month,

Year) Depending on

General parameter

17 (Date Format).

<Integer>
<Integer>
<Integer>

Range as follows:Month = 1 to 12

Day = 1 to 31

Year = 2000 to 2099

10

(0x0a)

Set time (Hour,

Minute, Second)

<Integer>
<Integer>
<Integer>

Range as follows Hour = 0 to 23

Minute = 0 to 59

Second = 0 to 59

11

(0x0b)

Zero Counter

<Counter>

<Byte>

Counter Range is 0-20

0 = All

12

(0x0c)

Zero Timer <Timer>

<Byte>

Timer Range is 0-20

0 = All