5.1.6.3.6 Install the Vent Tubing and Connect to Power

CAUTION

Never connect low and high pressure vent lines together.
3000 Micro GC will be damaged by back pressure.

1

Attach low pressure vent tubing to the low pressure vent on the heated sample
introduction accessory. See

Figure 5-34

.

For the Heated Vaporizer accessory
(PN G2819A-X/G2846A-X/G2858A-X),
use the clear tubing included in the kit

For the Heated Regulator accessory
(PN G2818A-X/G2845A-X/G2857A-X),
use 1/16 in. stainless steel tubing (not provided)

Figure 5-34 Attaching the vent tubing

2

Attach the copper high pressure vent tubing to the high pressure vent. For the
Heated Vaporizer accessory (PN G2819A-X/G2846A-X/G2858A-X),
use the 1/8 in. copper vent tubing included in the kit.

WARNING

If a hazardous sample is used, the sample exhaust must
be vented to a fume hood or other designated area for
hazardous waste disposal.

If desired, install a flow meter inline from the low pressure vent tubing
to check flow rates.

Attach low pressure vent tubing

G2819A-X on a 2-channel 3000 Micro GC

Attach high pressure vent tubing

Relief valve