TCD and columns turn off
(continued)

High carrier gas pressure

Ensure the carrier gas is supplying at least
78 PSI, and no more that 82 PSI.

CAUTION

Exceeding 82 PSI for the
carrier gas pressure may
result in damage to
3000 Micro GC.

Malfunctioning column or
injector

Check the flow coming out of the analytical
and reference vents on the back of the
instrument. If there is no flow on either/both
vents, contact

INFICON

.

Malfunctioning TCD

Contact

INFICON

.

Malfunctioning EPC valve

Contact

INFICON

.

Inlet/column temperature not
stabilizing

Instrument requires power
cycle

Restart 3000 Micro GC as instructed in

section 6.3.2, Restart, on page 6-8

.

Hardware problem

Contact

INFICON

.

Unable to download
method/error message

Instrument requires power
cycle

Restart 3000 Micro GC as instructed in

section 6.3.2, Restart, on page 6-8

.

Incorrect configuration

See

section 7.2, Connecting EZ IQ with

3000 Micro GC, on page 7-1

.

EZ IQ requires restart

Close EZ IQ and re-open. Restart the
computer if necessary.

Method is corrupt

Create a new method and reprogram
parameters.

Drivers are corrupt

Configure the instrument through EZ IQ
Config
.

D-Board error

Contact

INFICON

.

TCD detector 'fault'

Instrument requires power
cycle

Restart 3000 Micro GC as instructed in

section 6.3.2, Restart, on page 6-8

.

EZ IQ requires restart

Close EZ IQ and re-open. Restart the
computer if necessary.

Drivers are corrupt

Configure the instrument through EZ IQ
Config
.