4.2.3 Making Carrier Gas and Sample Inlet Connectors

3000 Micro GC is equipped with a 1/8 in. carrier gas Swagelok connector on the
back of the instrument. This section demonstrates how to assemble 1/8 in. carrier
gas tubing with the instrument.

The following procedure reviews how to make Swagelok connections. This
procedure explains how to connect tubing to a fitting, such as the sample inlet(s),
carrier gas inlet(s), or the gas supply tank.

Materials Needed

1/8 in. OD pre-cleaned copper or stainless steel tubing

1/8 in. Swagelok nuts, and front and back ferrules

Two 7/16 in. wrenches for 1/8 in. tubing

Two 5/16 in. wrenches for 1/16 in. tubing

Procedure

1

Attach a 1/8 in. Swagelok nut, back ferrule and front ferrule to the tubing as
shown in

Figure 4-1

.

Figure 4-1 Swagelok nut and ferrules setup on tubing

2

Clamp a female fitting in a bench vise.

3

Push the tubing into the female fitting.

Female fitting (1/8 in.)

Front ferrule (1/8 in.)

Back ferrule (1/8 in.)

Swagelok nut (1/8 in.)

Tubing (1/8 in.)