7.13.2 External Standard (ESTD) Reports

An ESTD Report contains information on retention time, area, and mole percent
concentration. It is not method specific. The same template will be applied to any
method that is opened

.

7.13.2.1 Customize an ESTD Report

An ESTD can be modified to specific needs.

The ESTD Report can be changed in a similar manner as the method report. Once
a change is made, it will apply to all methods.

Follow Steps 1 through 5 in

Importing a Template File, see section 7.13.1.1.2 on

page 7-44

.

7.13.2.2 Save a Customized ESTD Report

1

Once all desired information has been modified or added, right-click in the
Method Report window and select Export Report. See

Figure 7-60

.

Figure 7-60 Report editor - export report

2

The Save Method Custom Report As window will display.

3

Click External Standard.srp.