3000 Micro GC Operating Manual

6

Install the sample outlet tube into the output fitting on the Genie Filter
Assembly. See

Figure 5-3

. Tighten 1/2 turn past finger tight using a 1/4 in.

open-ended wrench, then loosen 1/4 turn, so that the tube is loose but will not
disengage.

Figure 5-3 Installing the sample outlet tube

7

Align the Genie Filter Assembly bracket horizontally the 3000 Micro GC face so
that the end of the outlet tubing is near the inlet fitting. Then, place the lower lip
of the mounting bracket under the bottom front panel as shown in

Figure 5-4

.

The edge of the 3000 Micro GC front molding fits into the groove in the bracket.

Figure 5-4 Installing the Genie filter assembly

Install sample outlet tube here

Venting tube

G2817A Shown

Sample inlet fitting

Align groove in bracket
with the edge of the
front molding