8.5.2.3 Replace the Battery
To replace the battery:
1
Shutdown the 3000 Micro GC, refer to
section 6.3.3, Shutdown, on page 6-8
.
2
Turn off the carrier gas.
3
Turn off the power.
4
Disconnect the power cord.
5
Remove the screws from the Portable 3000 Micro GC top cover.
6
Remove the top cover.
7
Turn Portable 3000 Micro GC over and remove the screws from the bottom
cover.
8
Remove the bottom cover.
9
Facing the front panel of Portable 3000 Micro GC, the battery compartment is
located on the top left side. See
. The panel covering the
compartment is held by two Torx screws on the outer edge of the panel.
Remove these screws and remove the panel.
Figure 8-40 Portable 3000 Micro GC battery compartment
Battery compartment
cover
Battery