7.14 EZ IQ Operation to Activate Remote Start/Cancel Function

To use the remote connector function in EZ IQ:

1

Open EZ IQ.

2

Click Method >> Instrument Setup to display the Instrument Setup window.

3

On the Trigger tab, select Type: External. See

Figure 7-63

.

Figure 7-63 Select external trigger type in instrument setup window

4

Save the method by clicking File >> Method >> Save.

5

Download the method by clicking Control >> Download Method.

6

Enter Sample ID and enter Number of reps. See

Figure 7-64

.

NOTE: When multiple sample injections are defined in the Single Run

Acquisition window, it is required that each run be triggered by a
REMOTE_START contact closure by an external system. Once a run
is submitted, it will not start until the trigger is activated. Refer to

section

5.2, The Remote Connector, on page 5-30

.

Figure 7-64 Enter multiple samples into single run acquisition window