11.4 Checkout Gases

11.4.1 Universal Checkout Gas

Universal Checkout Gas may be used during installation to verify the chemical
performance of 3000 Micro GC or GC module. See

Table 11-9

for the component

concentrations.

NOTE: Do not use the Universal Checkout Gas as a calibration standard for

specific applications. Different applications have different calibration
requirements and must be calibrated by an application-specific calibration
standard gas.

Table 11-8 Checkout gases

Description

Part Number

Universal Checkout Gas Cylinder

5183-4810

NGA Checkout Gas Cylinder

5184-3556

RGA Checkout Gas Cylinder

5184-3558

Regulator for Checkout Gas Cylinders

5184-3559

Table 11-9 The universal checkout gas

Components

Concentration (Mole %)

helium

0.10%

neon

0.05%

hydrogen

0.10%

oxygen

0.05%

nitrogen

0.10%

methane

balance

ethane

0.05%

ethylene

0.05%

carbon dioxide

0.05%

carbon monoxide

0.10%

acetylene

0.05%

propane

0.05%

methyl acetylene

0.05%

n-butane

0.05%

n-hexane

0.05%

n-heptane

0.05%