7 - 31

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

4

Click Next >> to move to the next detected peak. Click << Back to move to the
previously detected peak in the chromatogram.

NOTE: The current peak and total peaks in the chromatogram are displayed

on the right of the window.

Alternatively, if a specific peak must be identified, click on the peak. The
Retention time shown in the window will change to reflect the selected peak.
A blue Define Single Peak window will display. See

Figure 7-40

.

Figure 7-40 Blue define single peak window

The Define Single Peak window will display the retention time for the selected
peak.

5

When finished adding peaks to the peak table, click Done.

NOTE: For peaks that are not of interest, leave Peak Name blank and click

Next.

The parameters in the Define Single Peak window are described in

Table 7-3

.

Table 7-3 Parameters

Parameters

Description

Peak name

Enter the name of the compound in this field.

Conc level

Concentration Level 1 is displayed. Enter the amount of this
compound for this concentration level.

NOTE: To set up more than one level for this compound,

enter 2 for Conc. level and the amount for that
level.

Continue to enter level concentrations until completing the
number of calibration levels desired.)

Units

Enter the desired units to be used for results.