6.1 Overview

3000 Micro GC has an embedded web page for easy access to features such as
viewing configuration information, IP Address modification, and carrier gas
modification. The embedded web page is accessed via a web browser.

NOTE: 3000 Micro GC is not connected directly to the internet. 3000 Micro GC

has an embedded web server. When a computer and 3000 Micro GC are
connected via a crossover or a LAN cable, the web browser running on the
computer will be connected to the 3000 Micro GC web server.

6.1.1 Determine Basic 3000 Micro GC Configuration

The embedded web page home page contains basic information on the
configuration of 3000 Micro GC.

To access the basic configuration on the home page:

1

Ensure 3000 Micro GC is powered on.

2

Open a web browser on the computer. Microsoft Internet Explorer is
recommended.

3

Type in the IP address of the instrument into the address bar. See

Figure 6-1

.

3000 Micro GC comes with a default IP address of 10.1.1.101. Refer to

section

4.3.2, IP Address Configuration, on page 4-16

for instructions on how to set up

communication between the instrument and the computer.

Figure 6-1 Internet Explorer address bar