3 integration off, 3 add or edit integration events, 1 manually add or edit integration events – INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual
Page 151
7 - 23
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
7.10.2.3 Integration Off
The Integration Off
event turns off the integration of the chromatogram from the
Start Time to the Stop Time. This event is useful if there are regions of the
chromatogram that are not of interest.
7.10.3 Add or Edit Integration Events
There are two ways to add or edit an integration timed event to a method. One way
is to add or edit the integration events manually. The other is to use graphical
programming options.
7.10.3.1 Manually Add or Edit Integration Events
1
Select a channel to add or edit integration events. To change the channel,
select the appropriate channel from the drop-down list on the toolbar. See
.
Figure 7-26 Changing channels
2
Click Method >> Integration Events. The Integration Events window will
display the default integration events. See
Figure 7-27 Integration events window - default integration events
3
To add a new event, click on a blank field under Event in the table. Select the
desired integration event from the drop-down menu. See
.
Figure 7-28 Adding integration events