8.4.2 Replacing the 2 Micron Filter in the Heated Vaporizer

(PN G2819A-X/G2846A-X/G2858A-X)

The 2 micron filter assembly uses a replaceable cartridge filter (PN 0100-2034).
See

Figure 8-37

.

Figure 8-37 2 micron filter assembly

If the sample vessel is installed on the sample line, remove it as follows:

1

Close the stopcock on the sample vessel.

2

Turn the relief valve to Vent on the Heated Vaporizer and purge all residual
pressure remaining in the sample line.

3

Turn the relief valve to Sample.

4

Disconnect the sample vessel from the sample line at the quick disconnect.

5

Disassemble the trap halves and replace the filter assembly.