3000 Micro GC Operating Manual

3

Click Control >> Instrument Status to display the EZ IQ Instrument Status
screen. Wait until the method setpoints are all reached (i.e., the 3000 Micro GC
is Ready). Refer to

section 7.6.4, View Instrument Status, on page 7-18

.

4

Open the web browser and connect to the instrument via the embedded web
page. See

Chapter 6

.

5

Click

Utility.

6

Click Shutdown to shutdown the instrument.

7

After three minutes, set the front panel power switch to the off position.

CAUTION

Do not turn off the power without waiting at least three
minutes. 3000 Micro GC must update configuration files
before power is turned off. If power is turned off too soon,
the files could be corrupted, making 3000 Micro GC
inoperable.

8.1.3 Standby

Leave 3000 Micro GC power on with carrier gas flowing when 3000 Micro GC is
not in use. This enhances instrument performance and lifetime.