INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual
Page 215
8 - 33
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
2
Turn 3000 Micro GC on.
3
Wait three minutes until 3000 Micro GC has fully powered up.
4
Use EZ IQ software to load the desired method.
5
Power on the Heated Vaporizer.
6
Wait approximately five minutes for the Heated Vaporizer Assembly to stabilize
at operating temperature.
7
Make sure the sample vessel stopcock is closed and the relief valve is turned
fully to Sample.
8
Connect the sample vessel to the filter disconnect assembly on the Heated
Vaporizer. See
.
Figure 8-35 Sample vessel installed on a heated vaporizer (2-channel instrument shown)
9
Open the stopcock on the sample vessel.
10
Slowly turn the relief valve toward Vent until a small but steady flow is vented
and allow the sample to purge the line for approximately thirty seconds.
11
Turn the relief valve to Sample.
12
Use EZ IQ software to load the desired method and perform the analysis.
13
When the analysis is complete, close the stopcock on the sample vessel.
Sample vessel
Sample vessel stop-cock
Filter disconnect assembly
Quick disconnect fitting
Low pressure vent tubing
High pressure vent tubing
Relief valve