6.2.2 Check the Duty Cycle

1

The duty cycle should be approximately 30 to 40% for column head pressure
(CHP) and Delta P (if fixed volume).

2

Is the Duty Cycle between 30 to 40%?

2a

YES—Process with analysis.

2b

NO—If the duty cycle is > 50%, check the carrier gas inlet pressure. Replace
the carrier gas filter (PN 3150-0602). Recheck the Duty cycle. If it is still > 50%,
contact

INFICON

.

6.3 Utilities

The Utilities tab is on the 3000 Micro GC embedded web page. See

Figure 6-8

.

Figure 6-8 Utilities tab

6.3.1 Set Date/Time

In the Utilities tab, click Set Date/Time to set the correct date and time for the
3000 Micro GC.