5.1.6.3.2 User Supplied Parts
T20 Torx screwdriver
1/16 in. stainless steel tubing for low pressure venting
1/8 in. copper tubing for high pressure venting (rated for the maximum supply
pressure)
5.1.6.3.3 Prepare the 3000 Micro GC and Sample Introduction Accessory
1
Turn off 3000 Micro GC and allow the sample inlet fitting to cool.
WARNING
3000 Micro GC has inlets which can be heated to 140°C.
Touching the inlets at operating temperatures can result
in injury. Extreme care should be taken to avoid touching
the heated inlet surfaces.
2
Move the 3000 Micro GC handle fully away from the front cover.
3
Remove any fittings or hardware currently installed on the 3000 Micro GC
sample fitting.
4
Remove the upper left cover screw and loosen the two right-most screws so
that the heated sample introduction accessory mounting plate will slide
beneath them. See
.
Figure 5-26 Preparing to mount the accessory (2-channel 3000 Micro GC shown)
Remove
screw
2-channel 3000 Micro GC
Sample input fitting
Channel A
Loosen
screws